THYSTER

The THYSTER generator has been specifically designed to meet the nitrogen needs to supply a Microwave Plasma Atomic Emission Spectrometer (MP-AES) of the Agilent range for only sulfur component analysis. Nitrogen is produced by pressure swing absorption (PSA) to remove oxygen, carbon dioxide and water from compressed air. The model is available with an integral oil free compressor, and is extremely quiet in operation. The Generator is controlled using the latest in HMI touch screen technology to display the process in real time, inlet/outlet pressures.
  • Advantages
  • Characteristics
  • Specifications
  • Demo
  • Quote

Advantages

Improve analytical instruments performance
The production of a constant flow of gas improves the consistency of the analysis results and hence reproducibility.

Improve laboratory efficiency
The relatively high gas volumes required by MP-AES instruments make cylinder supply inappropriate for such applications. A constant, uninterrupted gas supply eliminates interruptions of analyses to change cylinders.

Improve economy
- Quick return on investment < 1 years
- No gas cylinder rental bottles, no price inflation

Improve safety
Nitrogen produced at low pressure and ambient temperature removes the hazards associated with high pressure cylinders and liquid Dewars.

Characteristics

  • Fully regenerative PSA technology: reduced risk of gas contamination and phthalate free
  • Complete Plug and Play system specifically designed for the Agilent MP-AES
  • Configuration for organics analysis :
    - N2 : 25 L/min at 5.5 bar (80 psi), purity > 99.7%
    - Dry Air : 36.5 L/min at 6 bar (87 psi) separate in 3 way for POP, EGCM and monochromator gas
  • Configuration for determination of sulfur :
    - N2 : 25 L/min at 5.5 bar (80 psi), purity > 99.7%
    - Dry Air : 36.5 L/min at 6 bar (87 psi) separate in 2 way for POP and EGCM optic purge
    - N2 : 10L/min at 5,5 bar (80 psi), purity > 99.95% to supply monochromator gas
  • With or without integral oil free air compressor with noise reduction technology
  • Quiet thanks to the Soundproofed compressor box and anti-vibration features
  • Auto start
  • Audible and alarm display with help menu and history log
  • Visual maintenance indication with history log
  • Outlet flow indicator
  • Energy saving Mode : Enables the compressor to switch off when nitrogen supply is not required
  • Remote PC monitoring via USB port.
  • Fit with wheels

Specifications

Models
Built-in Compressor
Flow rate
Purity and outlet pressure
Dimensions
(H x W x D)
Weight
(kg/lbs)
MP-AES Gas N2
POP Gas AIR
EGCM Gas AIR
Monochromat or
Purge gas
N2
AIR
Thyster/0
No
N2 10 L/min
> 99.95%
83 x 43 x 80 cm
(33" x 17" x 31")
88 / 194
5.5 bar
Thyster/1
Yes
(80 psi)
120 / 264.5









Technical information
Ambiant temperature range
5 - 25°C (41 - 77 °F)
Air inlet pressure (max)
8.5 bar (123 psi)
Air Inlet quality required (for unit without compressor)
Dewpoint : -40°C (-40°F)
Particulate : < 1 micron
Oil : < 0.01 mg/m3
Electrical supply
220 V ac/ 1 ph / 50 Hz
110 V ac / 1 ph / 50 - 60Hz
Inlet/outlet connections
G 1/4" (BSP) Female

Demo

 

Quote

 
KRSS Left Strip