N2-HP

The F-DGSi High Purity (HP) Nitrogen Generators are engineered to transform standard compressed air in to a safe regulated supply of 99.9995% pure nitrogen.
  • Advantages
  • Characteristics
  • Specifications
  • Demo
  • Quote

Advantages

Increased laboratory efficiency
A constant, uninterrupted gas supply of guaranteed purity eliminates interruptions of analyses to change cylinders and reduces the amount of instrument re-calibrations required.

Improved economy
Pure nitrogen gas produced as standard

Improved safety
Nitrogen produced at low pressure and ambient temperature removes the need for high pressure cylinders

Simple installation
Gas generators can be installed in the laboratory, on or under a bench, eliminating the need for long gas lines from cylinders secured elsewhere

Characteristics

  • Fully regenerative PSA technology; reduced risk of gas contamination and phthalate free
  • Generates Nitrogen on demand from external compressed air
  • Designed to run 24 hours a day
  • Smart control of the flow of inlet air, it is function of the real flow of nitrogen used by the application
  • Minimum maintenance with an annual filter change
  • Remote PC monitoring via USB/RS232.
  • Ultra fast start-up time and fast purity
  • Compact design, modular system

Specifications

Models

N2-600-HP.D

N2-1000-HP.D

N2 outlet
Flow rate (max) (cc/min)

600

1000

Outlet pressure (min)

1 bar (14.5 psi)

Outlet pressure (max)

Inlet pressure -2 bar (-29 psi)

Nitrogen purity

> 99.9995% (< 5 ppm O2 @ 7 bar (101 psi) inlet pressure)

O2 content

< 5 ppm

Air inlet
Air inlet quality requirement

Particulate : < 1 micron

Dewpoint : -40°C (-40°F)

Oil < 0.01 mg/m3

Flow rate (max)

15 L/min

18 L/min

Min supply pressure

7 bar (101 psi)

Max supply pressure

10 bar (145 psi)

Communication
LCD touch screen

Yes

Yes

RS232

X

RS485

X

USB

X

General Data
Supply rating

100 - 240 V ac / 50  - 60 Hz  2 A

Connection type

IEC320-C13

Installation power (max)

240 W (280 VA)

Fuse rating

4 A 240 V ac (5 x 20 mm)

Dimensions (W x H x D)

30 x 47 x 43 cm (11" x 18" x 17")

Net weight (kg/lbs)

< 25 / < 55

Connections
Outlet port

1/4" female

Inlet port

1/4" female

Operating conditions
Temperature

5 - 35°C (41-  95°F)

Humidity (max, non condensing)

80% at 25°C (77°F)

Noise

< 25 dB(A)

IP rating

IP20

Pollution degree rating

2 (with no aromatic compounds)

Altitude

< 2000 m

Demo

 

Quote

 
KRSS Left Strip